Bacterial Endotoxins Testing (BET)

Endotoxins testing, also known as the Bacterial Endotoxins Test (BET), is a quantitative microbiological assay used to determine the level of bacterial endotoxins present on medical devices, components, or materials. Endotoxins are lipopolysaccharide (LPS) fragments derived from the outer membrane of Gram-negative bacteria and can trigger severe inflammatory or febrile reactions if introduced into the human body.

Because endotoxins are heat-stable and not reliably inactivated by many sterilization processes, their control requires dedicated testing and preventive measures. Endotoxins testing is therefore essential for medical devices that come into contact with blood, cerebrospinal fluid, the cardiovascular system, or other critical body sites.

At Estereobato, endotoxins testing is performed using validated, cartridge-based detection workflows that provide rapid, quantitative results with high sensitivity and reproducibility. Our approach minimizes manual handling and procedural variability while meeting global regulatory expectations for medical device microbiological safety. We support manufacturers throughout the product lifecycle, from development and release testing to ongoing process monitoring and regulatory compliance.

The applications for our endotoxins testing

Endotoxins testing is applied across a wide range of medical device types and manufacturing scenarios. Typical application areas include:

  • Release testing of finished medical devices and components
  • Evaluation of product contact surfaces, coatings, and extractables
  • Supplier qualification and incoming material quality control
  • Monitoring of manufacturing process hygiene and water systems
  • Investigation of out-of-specification results or deviations
  • Support of risk assessments related to pyrogenic contamination
  • Provision of microbiological evidence for technical documentation and regulatory submissions

Endotoxins testing is particularly critical for devices intended for intravascular use, implantable devices, and products with prolonged or repeated patient contact.

Applicable standards and regulatory references

Endotoxins testing at Estereobato is conducted in accordance with internationally recognized pharmacopoeial and regulatory standards, including:

European Pharmacopoeia (Ph. Eur.) 2.6.14 – Bacterial Endotoxins
(Harmonised with USP <85> and JP 4.01)

ISO 11737 – Sterilization of health care products – Microbiological methods
(Provides a contextual framework for endotoxin control within sterility assurance strategies)

These standards define validated test principles, acceptance criteria, and system suitability requirements, ensuring results are acceptable to regulatory authorities worldwide.

Endotoxins tests in accreditated laboratory according to ISO 17025.
Tests performed in a state of the art laboratory with modern equipment.
Full service support in validating your test method.

Stages of endotoxins testing

Endotoxins testing is performed using validated, controlled, and traceable procedures. While specific conditions may vary depending on the product and intended use, a typical BET workflow includes the following stages:

1
Sample receipt and registration
Samples are received under controlled conditions and registered in a fully traceable laboratory system. Product details, test requirements, and applicable limits are reviewed prior to testing.
2
Sample extraction and dilution
Endotoxins are extracted from the device or material using pyrogen-free reagents. Extraction and dilution strategies are tailored to the product’s material composition, surface area, and geometry.
3
Interference evaluation
Potential inhibition or enhancement effects caused by the test article are assessed. This step ensures that the test system can accurately detect endotoxins in the presence of the product.
4
Quantitative endotoxin measurement
Endotoxin levels are measured using a validated kinetic chromogenic methodology. Cartridge-based systems provide precise, reproducible readings with minimal operator influence.
5
Calculation and evaluation of results
Results are calculated and evaluated against the applicable endotoxin limits, taking into account device category, patient contact type, and regulatory requirements.
6
Technical review and reporting
All data undergo technical review before issuance of a clear, standards-referenced analytical report suitable for regulatory use.

FAQs about our endotoxins testing

There are many questions about endotoxins testing, we have selected the most frequently asked questions. If you do not see your answer, please contact us.

What sensitivity range can endotoxins testing achieve?
Quantitative endotoxins assays typically achieve detection limits suitable for both low-level compliance thresholds and broader quality screening needs, for example down to 0.005 EU/mL, depending on the method and regulatory context.
Why is interference testing important?
Certain materials, coatings, or residual chemicals can inhibit or enhance the endotoxin signal, leading to inaccurate results. Interference testing ensures that reported values are valid, reliable, and defensible.
Can the method be validated for my specific device?
Yes. Product-specific method suitability and validation studies are standard practice to demonstrate that the chosen test conditions are appropriate for the device material and intended use.
Are endotoxins removed by sterilization?
Not necessarily. Endotoxins are heat-stable and may remain present after sterilization, which is why dedicated endotoxins control and testing are required.
Are results suitable for regulatory submissions?
Yes. Estereobato issues standards-referenced reports that are suitable for inclusion in technical documentation, risk management files, and regulatory submissions.